Clinical Accuracy Validation of the Moni-Patch Temperature Monitoring Patch System

Starting soon · Not applicable

Conditions studied: Body Temperature, Febrile

In brief

Moni-Patch Temperature Monitoring Patch System, a non-invasive, continuous temperature monitoring device designed to estimate core body temperature by applying a Sensor to the neck. This clinical accuracy validation study aims to verify the accuracy and repeatability of the Moni-Patch compared with an FDA-cleared continuous tympanic temperature monitoring device used as the clinical reference. Main objectives of this study are; To verify that the Moni-Patch and the reference device(tympanic) demonstrate acceptable agreement, as defined by Bland-Altman Analyses (bias within ±0.4°C and limits of agreement between -1.0°C and 1.0°C). To verify the consistency and repeatability of temperature measurements obtained from the Moni-Patch during continuous monitoring.

Key facts

Study ID
NCT07397078
Run by
Murata Vios, Inc.
People needed
105
Starts
2026-01-21
Expected to finish
2026-03-31
Last updated by the study team
2026-02-09

Who can join

Age: 7 and older. Sex: any. Healthy volunteers: accepted.