Belun SpO2 Accuracy Comparison to Arterial Blood CO-Oximetry

Completed

Conditions studied: Hypoxia

In brief

This study was a comparative, single-center, non-randomized study conducted to evaluate the SpO2 accuracy per standards and guidelines identified above for SpO2 accuracy for pulse oximetry equipment over the range of 70-100% SaO2 under non-motion conditions. Arterial blood sampling measured by functional SaO2 CO-Oximetry, was used as the basis for comparison. Testing was conducted under normal office environment conditions.

Key facts

Study ID
NCT04873947
Run by
Belun Technology Company Limited
People needed
10
Starts
2020-08-25
Expected to finish
2020-08-28
Last updated by the study team
2021-05-05

Who can join

Age: 18 and older, up to 50. Sex: any. Healthy volunteers: accepted.