BiOSENCY BORA Band SpO2 Validation Study

Completed · Not applicable

Conditions studied: Hypoxia

In brief

The purpose of this clinical study is to validate the SpO2 accuracy of the BiOSENCY BORA BAND™ wristband Pulse Oximeter during non-motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry. The end goal is to provide supporting documentation for the SpO2 accuracy validation of BORA BAND™ wristband Pulse Oximeter. It is required that the Accuracy Root Mean Square (ARMS) performance of the BORA BAND™ wristband Pulse Oximeter will meet a specification of 3.5 or better in non-motion conditions for the range of 70 - 100% SpO2 thereby demonstrating an acceptable SpO2 accuracy performance specification.

Key facts

Study ID
NCT03918018
Run by
Biosency
People needed
10
Starts
2019-02-19
Expected to finish
2019-02-21
Last updated by the study team
2019-04-17

Who can join

Age: 18 and older, up to 50. Sex: any. Healthy volunteers: accepted.