SpO2 Accuracy Comparison of Medline ReNewal Sensors to Arterial Blood CO-Oximetry

Completed · Not applicable

Conditions studied: Arterial Oxygen Saturation, Healthy

In brief

The purpose of this clinical study is to validate the SpO2 accuracy of the Medline ReNewal pulse oximetry sensors during non-motion conditions over the range of 70-100% SaO2 as compared to arterial blood samples assessed by CO-Oximetry. The end goal is to provide supporting documentation for the SpO2 accuracy validation of the ReNewal sensors. It is required that the Accuracy Root Mean Square (ARMS) performance of the ReNewal pulse oximetry sensors will meet a specification of 3 or better in non-motion conditions for the range of 70 - 100% SaO2 thereby demonstrating an acceptable SpO2 accuracy performance specification.

Key facts

Study ID
NCT03565679
Run by
Medline Industries
People needed
10
Starts
2018-05-09
Expected to finish
2018-05-10
Last updated by the study team
2019-06-18

Who can join

Age: 18 and older, up to 50. Sex: any. Healthy volunteers: accepted.