Evaluation of Blood Glucose Monitoring Systems With Blood Samples From Neonates
Completed · Not applicable
Conditions studied: Diabetes
In brief
The purpose of this study was to evaluate the performance of three Bayer Blood Glucose Monitoring Systems (BGMS) with neonatal blood. Blood samples used in this study were 'left-over samples'. The blood samples were from heel sticks of neonates, collected (into a tube) and sent to the laboratory. All meter Blood Glucose (BG) results were compared with capillary plasma results obtained with a reference laboratory glucose method (Cobas® 6000 instrument).
Key facts
- Study ID
- NCT01756274
- Run by
- Ascensia Diabetes Care
- People needed
- 162
- Starts
- 2012-12-01
- Expected to finish
- 2013-02-01
- Last updated by the study team
- 2017-02-15
Who can join
Age: any, up to 0. Sex: any. Healthy volunteers: accepted.
You may qualify if…
- Routine blood samples (heelstick) collected from neonates (less than 30 days of age).
- A portion of the samples (approximately 10%) will be from subjects <24 hours old.
- A portion of the samples (at least 10%) will be from subjects in the NICU.
- Sample blood volume must be sufficient to complete investigational testing in addition to prescribed testing.