Noninvasive Blood Pressure Monitor Evaluation Including Performance During Dysrhythmias
Completed · Not applicable
Conditions studied: Healthy
In brief
This protocol focuses on the evaluation of the GE noninvasive blood pressure parameter in subjects with dysrhythmias and normal sinus rhythm. The study follows the recommendations of the ANSI/AAMI SP10 Standard.
Key facts
- Study ID
- NCT00443001
- Run by
- GE Healthcare
- People needed
- 40
- Starts
- 2007-04-01
- Expected to finish
- 2007-07-01
- Last updated by the study team
- 2022-07-27
Who can join
Age: any. Sex: any. Healthy volunteers: accepted.
You may qualify if…
- signed informed consent
- a patient who is undergoing a specialized procedure such as a surgical procedure or cardiac catheterization, or during critical care monitoring in which a central arterial catheter is in place to determine systolic, diastolic, and/or mean arterial blood pressure values.
- ability to monitor ECG, NIBP and/or pulse oximetry
- presence of atrial fibrillation, atrial flutter, or normal sinus rhythm
You may not qualify if…
- any subject deemed too unstable, at the clinician's discretion, to participate in the study
- any subject with a cardiac or vascular malformation that results in an abnormal waveform inappropriate to be considered a reference standard
- any subject with a cardiac anomaly that would cause a disparity between aortic pressures and the periphery.
- presence of peripheral vascular disease in arms
- inability to perform a minimum of 4 fast flushes (two at the study beginning and two at the study end) on an arterial catheter or line in order to determine a frequency response. The volume of fluid will not exceed a total pre-specified by the Principal Investigator or designee.
- if the baseline arm to arm mean difference is > 10 mmHg, then the differences will be recorded and evaluated by the Principal Investigator for possible subject exclusion from the data analysis.
- subjects who, at the clinician's discretion, could not tolerate approximately 5 to 12, but possibly up to 25 repeated blood pressures on each arm
Where it is running
- GE Healthcare — Milwaukee, Wisconsin, United States